        <p>摘要：目的：统计与分析 DR 设备检测数据中的放射性（RA）性能情况，以期为采购设备及设备选择提供
依据。方法：收集本省三级综合（简称三级）、二级综合（简称二级）医院 DR 设备共 150 台的质控检测情况
及相关数据，按实际启用年限、设备国别、医院类型进行统计对比。结果：本组 150 台 DR 设备中，三级医院
为 100 台，其中进口设备占 80.00%（80/100）；二级医院为 50 台，其中进口设备占 60.00%（30/50）。统计 DR
设备实际启用年限显示，启用 5 年以上占 70.00%（105/150），启用 5 年以下占 30.00%（45/150）。结论：三级
医院中进口 DR 设备占比较大，且具有较长实际启用年限，且国产与进口相比在质量可靠性和稳定性上略有不
足。</p>