      <abstract>
        <p>目的：为解决传统手工室内质控流程复杂、耗时耗力的问题，本研究基于安图生化免疫流水线，对盘锦辽油宝石花医院检验科 2025年 1 月至 8 月的生化与免疫室内质控结果展开分析，验证自动质控的应用效果。方法：研究采用日本佳能生化分析仪、郑州安图化学发光分析仪等设备，以标源生物有限公司质控品为对象，通过 “每周一次加样密封、每日自动调取检测” 模式运行 8 个月，从质控品稳定性、检验效率、成本及工作量多维度评估。结果：所有检测项目的变异系数均低于室间质量评价标准允许值；生化与免疫检验项目 TAT 时间分别从 245 分钟、312 分钟缩短至 213 分钟、255 分钟；生化质控品用量从每月 4 瓶减至 3 瓶，免疫质控品用量从每月 6 管减至 3 管。结论：生化免疫流水线自动质控可保障结果稳定、提升检验效率并降低成本，为临床实验室质控自动化升级提供实践方案，但存在研究样本单一、故障分析不足等局限，未来需多中心研究与流程优化。</p>